High Accuracy Milliamp Loop Diagnostic Calibrator
PIECAL 434


Do More with the Evolutionary PIECAL 434!

•  Calibrate Milliamp Instruments
Source 0.000 to 24.000 mA & -25.00 to 125.00% of 4-20 mA
Read 0.000 to 24.000 mA & -25.00 to 125.00% of 4-20 mA
Read -99.99 to 99.99 V
Simulate 2-Wire Transmitter, Power & Read Transmitters
Menu Selectable 250 Ohm resistor for HART compatibility
•  Calibrate with Lab Accuracy
Accurate to ±(0.01% of Reading + 0.002 mA)
PIECAL 434 is more accurate than similar calibrators - within ±0.004 mA at 20 mA output!
•  Easty-to-Read Backlit Display
•  Locate Loop Current Leakages
Automatic indication of Loop Current and Leakage Current. Measure ground current leakage from faulty wiring, flooded conduit and corrosion bridges. Leakage typically appears as a Zero Shift and helps you decide if the transmitter needs calibration or if there is a problem in the loop wiring.
•  Faster Calibrations
Three output EZ-CHECK switch
2 speed EZ-DIAL
Stepping & Ramping
•  Includes Rubber Boot & Test Leads
•  Optional Accessories
AC Adaptor 120 V (020-0101); 240 V (020-0100)
NiMh Batteries & Charger 120 VAC/12V DC (020-0103)
 
Ground Leak Detection
If you find a loop where the transmitter is calibrated correctly but all the readings elsewhere in the loop have a fixed offset this is due to a Zero Shift. This zero shift is typically caused by some current in the loop bypassing the transmitter. This might be caused by ground faults, moisture or corrosion.
Turn on Ground Leak Detection and use the PIECAL 434 to power up the loop. Any current that isn't controlled by the transmitter or other current control element will be indicated as leakage on the calibrator's display.
 
Typical problem found with Loop Diagnostics
error1 pic    error2 pic
Here is a loop where a technician has just recalibrated the transmitter but the control room still sees a problem. The problem started just after a rainstorm.   Using a PIECAL 434 to power the loop the technician detects a leakage of 0.51 mA - the same offset seen in the control room. He walks the loop and opens a junction box releasing a stream of water. The loop is again in control.
 
 
Advantages of the PIECAL 434 over the Altek 434
PIE 434
•  Larger Display with Backlight
Easy viewing in dark locations
•  Easy selection of 250 Ω Hart Compatible Resistor
No disassembly to move internal jumper
•  Detects Loop Current Leakages
Measure ground current leakage from faulty wiring, flooded conduit and corrosion bridges. Leakage appears as a Zero Shift and helps you decide if the transmitter needs calibration or if there is some problem in the loop wiring.
•  More Resistant to Accidental Damage
Rubber Boot and Stronger Switches
•  Less Expensive Battery Changes
Four versus six "AA" batteries
•  Includes Certificate of Calibration
All PIECAL Calibrators come with a free signed and dated certificate. Altek's come only with an unsigned Statement of Calibration Practices with a certificate of calibration available for an extra charge.